A novel stent flow chamber system demonstrates reduced thrombogenicity of bioresorbable magnesium scaffolds | Scientific Reports

News

HomeHome / News / A novel stent flow chamber system demonstrates reduced thrombogenicity of bioresorbable magnesium scaffolds | Scientific Reports

Nov 05, 2024

A novel stent flow chamber system demonstrates reduced thrombogenicity of bioresorbable magnesium scaffolds | Scientific Reports

Scientific Reports volume 14, Article number: 26691 (2024) Cite this article Metrics details Coronary artery disease (CAD) is characterized by narrowing and subsequent blockade of coronary arteries,

Scientific Reports volume 14, Article number: 26691 (2024) Cite this article

Metrics details

Coronary artery disease (CAD) is characterized by narrowing and subsequent blockade of coronary arteries, and imposes a significant health and economic burden. Stent and scaffold devices are introduced in advanced CAD to improve vascular stability and restore blood flow. Although in vitro flow systems like the Chandler loop have been developed to enhance the understanding of interactions between device materials, their coatings, and vascular cells, imaging-based in vitro analysis of device performance is limited. In this study, we established a novel stent flow chamber system designed to assess the thrombogenicity of bioresorbable magnesium scaffold (RMS) and stent materials in vitro. Additionally, we compared the thrombogenicity – an important clinical parameter in stent performance – of the Magmaris-316 L stainless steel stent with its predecessors, Magmaris RMS and a prototype of the third-generation RMS (DREAMS 3G). Analysis of platelet adhesion and coverage of the different devices under flow conditions demonstrated that the Magmaris RMS exhibits reduced thrombogenicity compared to the Magmaris-316 L stainless steel stent. Moreover, thrombogenicity of the DREAMS 3G prototype, composed of BIOmag material, is further decreased compared to its predecessors. The observed reduction in thrombogenicity of the DREAMS 3G prototype in vitro suggests additional improvements in clinical safety and efficacy, highlighting its promise for treating CAD. Future research on this prototype may thus open avenues for analyzing other blood components and patient-derived endothelial cells. In line with the 3R principles, this approach may also help reduce the need for animal testing.

Cardiovascular diseases remain a leading cause of morbidity and mortality worldwide, taking an estimated 17.9 million lives each year according to the World Health Organization1. Coronary artery disease (CAD), a condition defined by the narrowing of blood vessels in the heart due to plaque deposits, poses a major health and economic burden to the society. The standard of care for advanced CAD is the introduction of stents at the vascular stenosis site, aimed to support the mechanical stability of the impaired vessel and restore blood flow2. Stent technology and materials have become increasingly sophisticated over the years, leading to reduced thrombogenicity and improved biocompatibility, and allow for a safer and more efficacious treatment. Nonetheless, stent failure due to restenosis remains a major clinical problem in the treatment of CAD.

While in vivo analysis of device performance includes a multifactorial analysis, a controlled in vitro environment can be essential to investigate cell type-specific and flow rate-dependent changes. These controlled environments enable specific analysis of hemodynamics, biomechanics, or effects on selected blood components3,4, and can enhance our understanding of cellular mechanisms regulated by different device materials and coatings. However, current in vitro platforms, such as the Chandler loop or flow loop5, do not allow for direct high-resolution microscopic visualization of device-cell interactions, e.g., platelet adhesion and aggregation on the device surface.

Recent advances in the field of interventional cardiology include the development of bioresorbable scaffolds (BRSs). In contrast to permanent stent material, BRSs provide a temporary scaffold and are resorbed within 1–36 months, dependent on the backbone material6. This promising technology therefore has the potential to overcome many of the safety concerns associated with permanent stent material, such as hypersensitivity reactions, a delayed onset of neointimal growth, neo-atherosclerosis, and an increased risk of late or very late stent thrombosis6. Although bioresorbable magnesium scaffolds (RMSs) such as the Magmaris scaffold show particularly very low thrombosis in patients7,8, a direct and standardized comparison between RMSs remains difficult up to date.

Here, we compare a Magmaris-316 L stainless steel stent (backbone design equivalent) with Magmaris RMS9,10, and a prototype of the third-generation RMS scaffold (DREAMS 3G) manufactured by Biotronik, Bülach, Switzerland.

Using a novel stent flow chamber system, we confirm that Magmaris RMS showed reduced thrombogenicity in comparison to Magmaris-316 L10. Furthermore, our findings indicate that the thrombogenicity of the novel DREAMS 3G scaffold was even further reduced, implying the sustained and potentially improved clinical safety and efficacy for therapeutic applications. Our findings provide valuable insights into the intricate dynamics of platelet interactions with different device material and could provide the foundation for further in vitro assessments of RMSs and their clinical applications.

Buffy coats from healthy donors were obtained from voluntary blood donations at the Institute of Transfusion Medicine at the University Medical Center Hamburg-Eppendorf, Germany. Written informed consent was obtained from each patient included in the study. The study protocol conforms to the ethical guidelines of the 1975 Declaration of Helsinki and the regional ethics committee approved the study (PV6039). For some buffy coats the requirement for informed consent was waived by the Central Ethics Committee, Hamburg, Germany given the discarded nature of the blood samples.

Blood bags containing about 60 ml of buffy coat from citrated whole blood were provided as described above. To obtain platelet-rich plasma (PRP) pooled buffy coats were centrifuged at 800xg for 5min at room temperature (RT) with minimal acceleration and deceleration to separate remaining leukocytes and erythrocytes. Platelet counts were determined using the Siemens Healthineers Advia 2100 system according to manufacturer’s instructions. PRP was adjusted with autologous platelet-free plasma (PFP) to 3-4 × 105 platelets/µl.

Two RMSs, DREAMS 3G9prototype and Magmaris RMS (DREAMS 2G)9,10, and a stainless-steel Magmaris-316 L (backbone design equivalent) were compared in this study (all from Biotronik AG, Switzerland). The DREAMS 3G prototype used in this study, has the identical scaffold design as the Magmaris RMS and Magmaris-316 L stents but is made from BIOmag material11. Therefore, a head-to-head comparison of those three devices that share an identical design (Fig. 1B) with no variations in strut thickness (all devices: 150 μm), width (all devices: 150 μm) and cell compartment size allows for specific analysis of the thrombogenic potential of the device materials. Device backbones are coated with biodegradable poly-L-lactide (PLLA) that is loaded with sirolimus as an immunosuppressive and anti-proliferative drug, which elutes 69.4% within 90 days after implantation9. The DREAMS 3G with the final design will be commercially available under the name Freesolve.

A novel in vitro flow chamber system to study thrombogenicity of scaffold material. (A) Schematic of scaffold embedding in side and front view. The open side up orientation of the half scaffold (with the cutting ends pointing upwards) in the channel slide was selected. (B) Cutting template of the stent and scaffold design to generate half stents and scaffolds using the indicated cutting edges. The stability of the scaffold is maintained via its connectors. (C) The scaffold embedded in the flow chamber is attached to the ibidi pump system, which is set up following manufacturer’s instructions. During the experiment fluidic units are placed inside a temperature-controlled CO2-incubator. (D) Micro-thrombotic events occurred at all scaffolds in the area of direct contact between the strut and glass coverslip. Individual flow experiments were conducted with PRP obtained from the same buffy coat for all scaffolds. Laminar shear stress of 10dyn/cm2 was applied for 2 h. Representative immunofluorescence of platelets attached to Magmaris-316 L, Magmaris and DREAMS 3G prototype (PT) scaffolds using an antibody against CD62P (green) and phalloidin (red, actin). Single channel images are depicted in grey. (E) Platelet accumulation and coverage of different scaffold materials can be reliably analyzed along the strut length under laminar flow. Individual flow experiments were conducted with PRP obtained from the same buffy coat for all scaffolds. Laminar shear stress of 10dyn/cm2 was applied for 2 h. Representative immunofluorescence of platelets attached to Magmaris-316 L, Magmaris and DREAMS 3G prototype scaffolds using an antibody against CD62P (green) and phalloidin (red, actin). Single channel images are depicted in grey. Scale bars: 50 μm (D, E).

Scaffold and stent devices were inflated under semi-sterile conditions using the Encore 26 Inflation device (Boston Scientific, MA, USA) according to the manufacturer’s protocol. Device backbones are designed with six crowns in each row and all adjacent rows are interconnected with two axial connectors. Devices were cut in half along a pre-defined area using a dissection microscope and sterile fine dissection instrument. With two rows of connectors per half device remaining, scaffold and stent stability was not compromised.

The embedding procedure into the sticky µ-Slide I Luer channel (channel height 0.8 mm, cat#80198, ibidi GmbH, Germany) was performed under semi-sterile conditions. The half scaffold or stent was oriented open-side-up with cut ends pointing upwards and the intact scaffold half pointing towards the sealing glass coverslip (cat#10814, ibidi GmbH, Germany). The glass coverslip was attached to the sticky slide including the embedded device and moderate pressure was applied to fully fit the device into the channel and to seal the slide airtight. All slides were additionally sealed with super glue to prevent PRP leakage and loss of vacuum.

The ibidi pump system (cat# 10902, ibidi GmbH, Germany) and the ibidi PumpControl software were employed to subject half scaffolds to controlled flow conditions. During the experiment fluidic units were placed inside a conventional cell culture incubator at 37 °C in a humidified atmosphere with 5% CO2. 13 ml of PRP were added to the reservoirs and the menisci of both sides were equilibrated. 10 ml fluid reservoirs, perfusion set (cat#10962, ibidi GmbH, Germany) and the channel slide were mounted according to the manufacturer’s protocol. Air bubbles were removed using a pre-defined ibidi PumpControl program and then arterial-like12 laminar flow (shear stress: 10dyn/cm2, plasma viscosity at 37 °C around 1.2–1.3 cPoise13 (0.012–0.013 dyn*s/cm2)) was applied for 2 h allowing platelets to recirculate.

Precise shear stress values have been determined by ibidi GmbH, Germany (https://ibidi.com/flow-accessories/118-perfusion-set.html) based on perfusion set and channel specific parameters, such as tube lengths and diameter, channel diameters, and work volumes.

After laminar flow exposure, slides were detached from the flow unit and washed twice with phosphate-buffered saline (PBS). Adherent platelets were fixed with 4% (w/v) paraformaldehyde for 45 min at RT. Cells were permeabilized with 0.5% (v/v) Triton-X-100 solution at RT for 5 min. Unspecific binding sites were blocked with 2% (w/v) BSA plus 0.3% (v/v) Triton-X-100 in PBS for 1 h at RT. Immunostaining was performed using a mouse anti-CD62P antibody (1:100, cat# ab6632, Abcam, UK) for 1 h at RT. After washing three times with 0.3% (v/v) Triton-X-100 in PBS, cells were incubated with fluorophore-labeled AffiniPure Fab Fragment anti-mouse 488 (cat# 115-547-185, Jackson ImmunoResearch, UK), and Alexa Fluor 594 Phalloidin (cat# A12381, Invitrogen, Germany) for 45 min at RT. Half scaffolds and stents were covered with PBS and z-stack images were taken within 2–3 days after conducting the experiment using a Leica SP8 confocal microscope with a HC PL APO 63x/1.4 Oil CS2 objective and the Leica LAS X SP8 software.

Quantifications were done using Fiji ImageJ. Actin pixel intensities (integrated density) were measured from maximum intensity projection images of acquired Z-stacks. To quantify category 2 platelets, spreading platelets were manually counted from 90 μm x 90 μm confocal z-stack images (8100 μm2).

As micro-thrombotic events were observed in all half devices within the region of direct contact between the strut and the glass coverslip, we selectively quantified device-blood component interactions along the strut length, while avoiding areas of artificially perturbed flow. Therefore, quantifications were not performed in a total blinded manner. Live videos of platelet flow were taken using a DS-Fi3 camera on the Nikon Eclipse Ts2R.

Statistical analysis. GraphPad Prism 9 was used for graphic representation and statistical data analysis. We used a two-tailed unpaired Student’s t-test to compare two means with equal variance. Differences were considered statistically significant when p < 0.05.

To analyze the thrombogenicity of coronary scaffolds in vitro, scaffold material was embedded in commercially available µ-Slide I Luer channel mimicking a blood vessel lumen (Fig. 1A). To avoid scaffold strut collapse of high-pressure inflated 3 mm diameter scaffolds in the µ-Slide, scaffolds were cut in half along a pre-defined line to that prevents scaffold structure disintegration (Fig. 1B). With two rows of connectors per half scaffold remaining, scaffold stability was not compromised. To insert a half scaffold into the channel slide, two orientations were tested: (i) The open-side-up orientation and (ii) open-side-down with cut ends pointing upwards or downwards, respectively. Half scaffolds embedded in the open-side-down orientation induce microfractures of the sealing coverslip caused by the sharp end of the cutting edges (data not shown). In contrast, the open-side-up scaffold orientation was technically easy to mount (Fig. 1A). Furthermore, the open-side-up orientation of a half scaffold allowed for convenient imaging of the entire half scaffold within the range of the objective’s working distance. To allow for long-term exposure of blood components to the half scaffold under physiological conditions, the channel slide containing the half scaffold was subsequently placed in a conventional cell culture incubator (Fig. 1C).

In order to test for thrombogenic events on different stent and scaffold surface material, freshly isolated platelet-rich plasma (PRP) from buffy coats was perfused in flow chamber systems with Magmaris-316 L, Magmaris RMS and DREAMS 3G prototype devices. Devices were exposed to a laminar platelet flow of 10dyn/cm2 for 2 h. High-resolution immunofluorescence imaging of the total scaffold and stent strut areas revealed micro-thrombotic events on all half devices within the region of direct contact between the strut crown and the glass coverslip (Fig. 1D), which were likely caused by disturbed flow in that particular area, irrespective of the scaffold material’s nature (Supplemental video 1). This specific area was therefore excluded from consecutive analyses. Notably, along the strut length, where laminar-like flow was obtained (Supplemental video 2), platelet adhesion and coverage were strongly reduced on DREAMS 3G prototype and Magmaris RMS half scaffolds compared to the control Magmaris-316 L device (Fig. 1E).

Next, we aimed to evaluate in vitro thrombogenicity of the novel DREAMS 3G prototype scaffold made of the BIOmag material, which is the next-generation development of its predecessor, the Magmaris RMS scaffold. We first compared platelet coverage of the DREAMS 3G prototype half scaffold to the backbone equivalent Magmaris-316 L. The devices were exposed to PRP samples from the same donor and exposed to laminar flow of 10dyn/cm2 for 2 h. We found that platelet coverage of DREAMS 3G prototype half scaffolds was significantly reduced by 77% when compared to Magmaris-316 L half device (Fig. 2A, B). Next, we sought to compare in vitro thrombogenicity of the DREAMS 3G prototype scaffold directly to its predecessor scaffold Magmaris RMS. Procoagulant surface-mediated platelet activation proceeds via rolling, initial tethering, firm adhesion and spreading14, Fig. 2C). We categorized platelet adhesion into category 1 (rolling and hemisphere-shaped) and category 2 (spreading) (Fig. 2C). To investigate scaffold-driven platelet activation we focused on firm adhesion and spreading events (category 2) to Magmaris RMS and DREAMS 3G prototype scaffolds and performed high-magnification immunofluorescence imaging. We found that firm adhesion and spreading of platelets to the BIOmag material of the DREAMS 3G scaffold was strongly reduced by 84% compared to the Magmaris RMS material (Fig. 2D), confirming its reduced thrombogenicity.

Thrombogenicity of the DREAMS 3G prototype made from BIOmag material is significantly diminished. (A) Individual flow experiments were conducted with PRP obtained from the same buffy coat for all scaffolds. Laminar shear stress of 10dyn/cm2 was applied for 2 h. Representative immunofluorescence of platelets attached to Magmaris-316 L and DREAMS 3G prototype (PT) scaffolds using an antibody against CD62P (green) and phalloidin (red, actin). (B) Quantification of platelet coverage (n = 3–6 images per scaffold from three independent experiments). Data represent mean ± s.e.m. p value, Two-tailed unpaired Student’s t-test. (C) Representative immunofluorescence of platelets attached to scaffold material in a reversible manner (category 1: rolling, ball-shaped and hemisphere-shaped) and in an irreversible manner (category 2: spreading platelet). (D) Irreversible platelet adhesion to the DREAMS 3G prototype scaffold is significantly diminished. Individual flow experiments were conducted with PRP obtained from the same buffy coat for all scaffolds. Laminar shear stress of 10dyn/cm2 was applied for 2 h. Quantification of irreversibly attached platelets (category 2) (n = 4 images per scaffold from two independent experiments). Data represent mean ± s.e.m. p value, Two-tailed unpaired Student’s t-test. Scale bars: 50 μm (A), 5 μm (C).

There is a demand for in vitro stent systems that can function as platforms to study scaffold and stent devices in a controlled environment. These systems provide valuable insights that complement in vivo studies, guiding the development of innovative devices. Simultaneously, they contribute to efforts aimed at minimizing animal experimentation in accordance with the principles of Replacement, Reduction, and Refinement (3R)15. Here, we established a novel and easy-to-use in vitro stent flow chamber system to study thrombogenicity of different stent and scaffold material. In a proof-of-concept study, our system revealed reduced thrombogenicity of the novel DREAMS 3G prototype exhibited strongly reduced in vitro thrombogenicity compared to its predecessor scaffold Magmaris RMS and the Magmaris-316 L stent backbone equivalent.

While contemporary drug eluting stents have a 2-year thrombosis rate of approximately 0.6-1%, RMS devices consistently showed very low 2-year scaffold thrombosis rates in clinical trials16,17. For example, improved clinical safety and reduced thrombogenicity of the Magmaris scaffold has been reported in a pooled analysis of BIOSOLVE-II and -III7with no scaffold thrombosis up to 36 months, and 0.8% (or 0.4% without early antiplatelet or anticoagulant interruption) scaffold thrombosis after 25 months in the BIOSOLVE -IV full cohort study8. Furthermore, the final DREAMS 3G design, tested in the BIOMAG-I study18, also showed 0% definite or probable scaffold thrombosis, confirming the low thrombogenic potential of magnesium based resorbable scaffolds in general.

A previous study using a porcine arteriovenous shunt model could further demonstrate that the Magmaris RMS exhibits less thrombogenicity and inflammatory cell deposition compared not only with the Magmaris-316 L backbone equivalent10but also compared to Orsiro, a contemporary ultrathin drug eluting stent with extensive clinical evidence19,20. Notably, our in vitro analysis reveals that the BIOmag material in the DREAMS 3G prototype significantly outperforms its predecessor, Magmaris RMS, by exhibiting reduced platelet coverage and spreading. Our findings are in line with a first-in-human study demonstrating that at the end of the resorption period the final DREAMS 3G device is clinically safe and effective11. Furthermore, the BIOmag material showed a significantly lower scaffold discontinuity when compared to Magmaris RMS material9, likely providing a more efficient and longer scaffolding time.

While the in vitro stent flow chamber system introduces fluid flow and shear stress, it lacks other mechanical forces of the physiological in vivo microenvironment such as extracellular matrix stiffness or stretch21,22. Micro-thrombotic events were observed in all half devices within the region of direct contact between the strut and the glass coverslip. This observation indicates that the occurrence of micro-thrombotic events in this area is artificially induced, resulting from disturbed flow, irrespective of the scaffold material’s nature. The in vitro stent flow chamber system therefore should only be utilized to investigate blood component interactions along the strut length, while avoiding areas of artificially perturbed flow. Future flow systems should implement a physiological vessel wall microenvironment, including hydrogel embedding and extracellular matrix components. This approach would also create a natural vessel-shaped anatomy around the devices.

Future applications employing the novel flow chamber system could include the study of device endothelialization23,24 and live imaging of blood component and endothelial cell interaction with device material. Furthermore, the in vitro stent flow chamber system has the potential to advance personalized medicine, e.g., by optimizing device material and drug elution tailored to patient needs or by analyzing and mitigating the individual’s risk of thrombosis and restenosis.

All data is provided within the manuscript or supplementary information files. Raw data are available from the corresponding author upon request.

Organization, W. H. Cardiovascular diseases (CVDs). (2021).

Scafa Udriște, A., Niculescu, A. G., Grumezescu, A. M. & Bădilă, E. Cardiovascular Stents: A Review of Past, Current, and Emerging Devices. Materials (Basel) 14. (2021). https://doi.org/10.3390/ma14102498

Filipovic, N. et al. In vitro and in silico testing of partially and fully bioresorbable vascular scaffold. J. Biomech. 115, 110158. https://doi.org/10.1016/j.jbiomech.2020.110158 (2021).

Article PubMed Google Scholar

Abbasnezhad, N., Zirak, N., Champmartin, S., Shirinbayan, M. & Bakir, F. An Overview of In Vitro Drug Release Methods for Drug-Eluting Stents. Polymers (Basel) 14. (2022). https://doi.org/10.3390/polym14132751

Link, A. et al. In vitro investigation of an intracranial flow diverter with a fibrin-based, hemostasis mimicking, nanocoating. Biomed. Mater. 16, 015026. https://doi.org/10.1088/1748-605X/abc8d3 (2020).

Article PubMed Google Scholar

Seguchi, M. et al. Challenges of the newer generation of resorbable magnesium scaffolds: lessons from failure mechanisms of the past generation. J. Cardiol. 81, 179–188. https://doi.org/10.1016/j.jjcc.2022.09.003 (2023).

Article PubMed Google Scholar

Haude, M. et al. Sustained safety and performance of the second-generation sirolimus-eluting Absorbable Metal Scaffold: pooled outcomes of the BIOSOLVE-II and -III trials at 3 years. Cardiovasc. Revasc Med. 21, 1150–1154. https://doi.org/10.1016/j.carrev.2020.04.006 (2020).

Article PubMed Google Scholar

Starmer, G. et al. TCT-81 safety and efficacy of the Resorbable Magnesium Scaffold magmaris in a Real-World setting: 36-Month outcomes of the BIOSOLVE-IV Registry. J. Am. Coll. Cardiol. 82, B32–B33. https://doi.org/10.1016/j.jacc.2023.09.086 (2023).

Article Google Scholar

Seguchi, M. et al. Preclinical evaluation of the degradation kinetics of third-generation resorbable magnesium scaffolds. EuroIntervention 19, e167-e175. (2023). https://doi.org/10.4244/eij-d-22-00718

Lipinski, M. J. et al. Comparison of acute thrombogenicity for magnesium versus stainless steel stents in a porcine arteriovenous shunt model. EuroIntervention. 14, 1420–1427. https://doi.org/10.4244/eij-d-17-00958 (2019).

Article PubMed Google Scholar

Haude, M. et al. Safety and performance of the third-generation drug-eluting resorbable coronary magnesium scaffold system in the treatment of subjects with de novo coronary artery lesions: 6-month results of the prospective, multicenter BIOMAG-I first-in-human study. eClinicalMedicine. 59, 101940. https://doi.org/10.1016/j.eclinm.2023.101940 (2023).

Article PubMed PubMed Central Google Scholar

Cheng, C. et al. Large variations in absolute wall shear stress levels within one species and between species. Atherosclerosis. 195, 225–235. https://doi.org/10.1016/j.atherosclerosis.2006.11.019 (2007).

Article PubMed Google Scholar

Vaupel, P. et al. From Localized Mild Hyperthermia to Improved Tumor Oxygenation: Physiological Mechanisms Critically Involved in Oncologic Thermo-Radio-Immunotherapy. Cancers 15, 1394. (2023).

Kuwahara, M. et al. Platelet Shape Changes and Adhesion Under High Shear Flow. Thromb. Vascular Biology 22, 329–334. doi:https://doi.org/10.1161/hq0202.104122. (2002).

Article Google Scholar

Hubrecht, R. C. & Carter, E. The 3Rs and Humane Experimental Technique: Implementing Change. Animals (Basel) 9. (2019). https://doi.org/10.3390/ani9100754

Kok, M. M. et al. Two-year clinical outcome of all-comers treated with three highly dissimilar contemporary coronary drug-eluting stents in the randomised BIO-RESORT trial. EuroIntervention. 14, 915–923. https://doi.org/10.4244/eij-d-18-00336 (2018).

Article PubMed Google Scholar

Swedeheart Swedeheart Annual Report https:// (2022). www.ucr.uu.se/swedeheart/dokument-sh/arsrapporter-sh/arsrapport-2022/01-swedeheart-annual-report-2022-english

Haude, M. et al. A new resorbable magnesium scaffold for de novo coronary lesions (DREAMS 3): one-year results of the BIOMAG-I first-in-human study. EuroIntervention. 19, e414–e422. https://doi.org/10.4244/eij-d-23-00326 (2023).

Article PubMed PubMed Central Google Scholar

Forrestal, B. J., Case, B. C., Yerasi, C., Garcia-Garcia, H. M. & Waksman, R. The orsiro ultrathin, bioresorbable-polymer sirolimus-eluting stent: a review of current evidence. Cardiovasc. Revascularization Med. 21, 540–548. https://doi.org/10.1016/j.carrev.2019.12.039 (2020).

Article Google Scholar

Waksman, R. et al. Comparison of Acute Thrombogenicity for metallic and polymeric Bioabsorbable scaffolds: Magmaris Versus absorb in a Porcine Arteriovenous Shunt Model. Circ. Cardiovasc. Interv. 10 https://doi.org/10.1161/circinterventions.116.004762 (2017).

Gordon, E., Schimmel, L. & Frye, M. The Importance of Mechanical Forces for in vitro endothelial cell Biology. Front. Physiol. 11, 684. https://doi.org/10.3389/fphys.2020.00684 (2020).

Article PubMed PubMed Central Google Scholar

Carlantoni, C. et al. The phosphodiesterase 2A controls lymphatic junctional maturation via cGMP-dependent notch signaling. Dev. Cell. https://doi.org/10.1016/j.devcel.2023.12.002 (2023).

Article PubMed Google Scholar

Habib, A. & Finn, A. V. Endothelialization of drug eluting stents and its impact on dual anti-platelet therapy duration. Pharmacol. Res. 93, 22–27. https://doi.org/10.1016/j.phrs.2014.12.003 (2015).

Article PubMed Google Scholar

Van Belle, E. et al. Stent Endothelialization Circulation 95, 438–448. https://doi.org/10.1161/01.CIR.95.2.438. (1997).

Article ADS PubMed Google Scholar

Download references

We thank the Institute of Transfusion Medicine at the University Medical Center Hamburg, Germany, for obtaining healthy control plasma samples. This study was supported by Biotronik AG/Cortronik funds and the German Research Foundation (DFG) grant FR 4239/1-1 (to M.F.). We acknowledge financial support from the Open Access Publication Fund of UKE - Universitätsklinikum Hamburg-Eppendorf.

Open Access funding enabled and organized by Projekt DEAL.

Monja Müller and Lars Ludwig contributed equally.

Institute of Clinical Chemistry and Laboratory Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany

Monja Müller, Lars Ludwig, Hanna Englert, May Cathleen Müller, Sandra A. Hemkemeyer, Manu Beerens, Reiner K. Mailer, Thomas Renné & Maike Frye

German Centre of Cardiovascular Research (DZHK), Partner Site Hamburg, Luebeck, Kiel, Hamburg, Germany

Katharina A. Riedl, Sandra A. Hemkemeyer, Manu Beerens & Maike Frye

Department of Cardiology, University Heart and Vascular Center, University Medical Center Hamburg-Eppendorf, Hamburg, Germany

Katharina A. Riedl

Center for Interdisciplinary Cardiac Imaging, University Medical Center Hamburg-Eppendorf, Hamburg, Germany

Katharina A. Riedl

Center for Thrombosis and Hemostasis (CTH), Johannes Gutenberg University Medical Center, Mainz, Germany

Thomas Renné

Irish Centre for Vascular Biology, School of Pharmacy and Biomolecular Sciences, Royal College of Surgeons in Ireland, Dublin, Ireland

Thomas Renné

Biotronik AG, Bülach, Switzerland

Sabine Lang & Philine Baumann-Zumstein

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

Monja Müller: Investigation, Validation, Formal Analysis, Writing - Review & Editing; Lars Ludwig: Investigation, Validation, Formal Analysis, Writing - Review & Editing; Hanna Englert: Investigation, Writing - Review & Editing; Katharina A. Riedl: Writing - Review & Editing; May C. Müller: Writing - Review & Editing; Sandra A. Hemkemeyer: Writing - Review & Editing; Manu Beerens: Formal Analysis, Writing - Review & Editing; Reiner K. Mailer: Formal Analysis, Writing - Review & Editing; Thomas Renné: Formal Analysis, Writing - Review & Editing; Sabine Lang: Conceptualization; Resources; Philine Baumann-Zumstein: Conceptualization; Resources, Writing - Review & Editing; Maike Frye: Conceptualization, Methodology, Validation, Formal analysis, Investigation, Resources, Data Curation, Writing - Original Draft, Writing - Review & Editing, Visualization, Supervision, Funding acquisition.

Correspondence to Maike Frye.

S.L. and P.BZ. are employees of Biotronik AG. M.F. reports a relationship with Biotronik AG/Cortronik GmbH that includes: consulting or advisory and funding of grants. M.M., L.L., M.C.M, H.E., K.A.R., S.A.H., M.B., R.K.M., and T.R. have no conflict of interest to disclose.

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Below is the link to the electronic supplementary material.

Supplementary Material 1

Supplementary Material 2

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

Müller, M., Ludwig, L., Englert, H. et al. A novel stent flow chamber system demonstrates reduced thrombogenicity of bioresorbable magnesium scaffolds. Sci Rep 14, 26691 (2024). https://doi.org/10.1038/s41598-024-77266-0

Download citation

Received: 24 June 2024

Accepted: 21 October 2024

Published: 04 November 2024

DOI: https://doi.org/10.1038/s41598-024-77266-0

Anyone you share the following link with will be able to read this content:

Sorry, a shareable link is not currently available for this article.

Provided by the Springer Nature SharedIt content-sharing initiative